HMS Oxy Evo
Electromedical device with oxygen concentrator, ozone generator, electroporation and electrostimulation
- Technologies
- 4 in 1
- O₂ flow
- 1–6 L/min
- O₂ concentration
- 36–93 %
- Ozone
- 0.5–7.75 µg/mL
- Electroporation
- 120 V · 100–300 Hz
- Display
- 15" touch
Intended use
What it is for and who uses it
HMS OXY EVO is an active therapeutic electromedical device implementing the following technologies: Oxygen concentrator to supply oxygen for oxygen therapy; Ozone generator to supply ozone for ozone therapy; Electroporation for the transdermal inoculation of therapeutic active ingredients; Electrostimulation for vaginal problems relating to the pelvic floor musculature and urinary incontinence. In addition, HMS OXY EVO employs the Electroporation technologies also for aesthetic applications outside the scope of Directive 93/42/EEC as amended, namely for the transdermal delivery of cosmetic substances.
Qualified operators
The intended use of the appliance is that of a "medical device", used exclusively by personnel qualified in accordance with the law for applications in the fields of physiotherapy, rehabilitation, pain therapy and aesthetic medicine. Specific training is required for correct use of the product.
Treatments and areas
- Oxygen therapy
- Ozone therapy
- Transdermal inoculation of therapeutic active ingredients
- Vaginal problems of the pelvic floor musculature
- Urinary incontinence
Contraindications
- Oxygen therapy: contraindicated in the case of newborns; presence of obstructive sleep apnoea; severe respiratory acidosis; smoking. Depending on the specific situation, however, the physician may decide to override these contraindications, for example in the presence of severe distress or respiratory arrest.
- Ozone therapy: contraindicated in the case of pregnant women; patients suffering from hyperthyroidism; patients suffering from diseases of the cardiovascular system; patients with respiratory diseases; patients with blood disorders; patients with blood coagulation defects.
- Electroporation and electrostimulation: contraindicated for wearers of active implants such as pacemakers, insulin pumps, pericardial stimulators, hearing aids or others; for women who are pregnant and/or breastfeeding; for wearers of metal prostheses and osteosynthesis devices (including dental prostheses) close to the area to be treated; for subjects with ongoing inflammatory processes, skin lesions, neoplasms; do not treat subjects with sensitive skin, if necessary limit to very low output powers; during menstruation (unless prescribed by a physician); apply only to intact skin; in subjects with multiple sclerosis (owing to spastic paralysis and sensory disorders); in persons with Parkinson's disease (owing to rigidity from increased muscle tone and tremor); in the presence of febrile illness; in cases of epilepsy; in cases of thrombophlebitis; in cases of heart disease. Avoid applying the electrodes on the carotid sinuses, so as not to stimulate the vagovagal reflex, and around the head.
- Electrostimulation, warnings: use in close proximity (e.g. 1 m) to short-wave or microwave therapy may produce instability in the output parameters; applying electrodes near the thorax may increase the risk of cardiac fibrillation; given the current density on the electrodes (greater than 2 mA/cm²), maximum attention by the operator is recommended; do not apply stimulation across the head, directly on the eyes, on the mouth, on the front of the neck (especially on the carotid sinus) or from electrodes positioned on the chest and upper back or across the heart.
- In case of doubt, do not use without first consulting the opinion of the specialist physician.
List taken from the instructions for use. Read the warnings and instructions carefully before use.
Technology
HMS Oxy Evo is an active therapeutic electromedical device that implements in a single appliance an oxygen concentrator, an ozone generator and the electroporation and electrostimulation technologies. It is compact (23.4 × 23.0 × 13.8 cm, 7 kg) and is managed from a 15-inch colour touch display, from which the operating screen, the parameters and the start, pause and stop of the treatment are selected. The panel carries the O₂/O₃ outlet and the connection for electroporation.
Oxygen concentrator
The concentrator supplies oxygen for oxygen therapy with a flow adjustable from 1 to 6 L/min; the concentration depends on the selected flow: 93 % at 1 L/min, 66 % at 2 L/min, 56 % at 3 L/min, 46 % at 4 L/min, 41 % at 5 L/min and 36 % at 6 L/min. The output pressure is 20–30 kPa, with a Ø 6/4 mm connection. Delivery is continuous and the concentration level can be adjusted from the display even during treatment.
Ozone generator
The generator supplies ozone for ozone therapy with concentrations defined in relation to the flow: 7.75 µg/mL at 1 L/min, 2.75 µg/mL at 2 L/min, 1.556 µg/mL at 3 L/min, 0.958 µg/mL at 4 L/min, 0.683 µg/mL at 5 L/min and 0.5 µg/mL at 6 L/min. The device provides for connecting, via suitable tubing, the various applied parts that the clinician intends to use for oxygen therapy and ozone therapy treatments. The manual notes that ozone is toxic if inhaled.
Electroporation
Electroporation is provided for the transdermal inoculation of therapeutic active ingredients and, in the aesthetic field outside the scope of Directive 93/42/EEC, for the transdermal delivery of cosmetic substances. The section delivers up to 120 V @ 1000 ohm with an emission frequency from 100 to 300 Hz; depth of action, body area, intensity (1–100 %) and time are set from the display. The equipment includes the HMS.MCEP applicator, the 80×40 synthetic chamois electrode, the cable kit and the 40-60-100 elastic straps.
Electrostimulation
Electrostimulation is intended for vaginal problems relating to the pelvic floor musculature and urinary incontinence and uses the supplied HMS.MVRFEP vaginal probe. Maximum voltage 25 V @ 500 ohm, emission frequency 100–300 Hz, intensity adjustable from 1 to 100 %. For the handpieces with electrodes the manual prescribes a sufficient and even layer of CE-marked gel.
Handpieces and standard equipment
What comes in the box
Consumables: CE-marked gel for the handpieces with electrodes, Tubing and applied parts for O₂/O₃ at the clinician's choice, Active ingredients for electroporation
Technical specifications
Technical specifications
| Power supply | 110–230 V AC, 50–60 Hz |
|---|---|
| Max power consumption | 60 VA |
| Electrical safety class | II BF |
| IP rating | IP40 |
| External protection | Protection fuses 2 × 1.6 A |
| Noise | 50 dB(A) |
| Oxygen flow | 1 ÷ 6 L/min |
|---|---|
| Oxygen concentration | 36 % @ 6 L/min · 41 % @ 5 L/min · 46 % @ 4 L/min · 56 % @ 3 L/min · 66 % @ 2 L/min · 93 % @ 1 L/min |
| Output pressure | 20 ÷ 30 kPa |
| Output connection | Outer Ø 6 mm, inner Ø 4 mm |
| Ozone concentration | 0.5 µg/mL @ 6 L/min · 0.683 µg/mL @ 5 L/min · 0.958 µg/mL @ 4 L/min · 1.556 µg/mL @ 3 L/min · 2.75 µg/mL @ 2 L/min · 7.75 µg/mL @ 1 L/min |
|---|---|
| Output connection | Outer Ø 6 mm, inner Ø 4 mm |
| Max output voltage | 120 V @ 1000 ohm |
|---|---|
| Emission frequency | 100 ÷ 300 Hz |
| Intensity | 1 ÷ 100 % (1 % steps) |
| Parameters | Depth of action, body sector/area, intensity, time |
| Max output voltage | 25 V @ 500 ohm |
|---|---|
| Emission frequency | 100 ÷ 300 Hz |
| Intensity | 1 ÷ 100 % (1 % steps) |
| Display | 15" colour touch screen |
|---|---|
| Dimensions (H × W × D) | 23.4 × 23.0 × 13.8 cm |
| Weight | 7 kg |
| Operating temperature | 10 ÷ 40 °C |
| Operating humidity | 30 ÷ 75 % non-condensing |
| Operating pressure | 700 ÷ 1060 hPa |
| Storage and transport | −10 ÷ 60 °C · humidity 10 ÷ 93 % · 700 ÷ 1060 hPa |
| Output parameter tolerance | 20 % |
Output parameter tolerance: 20% (data from the instructions for use).
Compliance and certifications
Medical device · CE marking with Notified Body
- Notified Body
- 1370
- Manufacturer
- CTM Tech S.r.l.s., Terni (Italia)
- EUDAMED SRN
- IT-MF-000032025
- Quality system
- ISO 13485 · AQSCert
- IFU
- Rev. 01 · 2022-07-26
Documentation
Training and service
Training included at delivery, direct service from our Terni workshop, replacement unit in case of extended downtime, CEI EN 62353 electrical checks.
Also available for rental or leasing → options
Healthcare professionals area
HMS Oxy Evo – information for healthcare professionals
Content reserved for healthcare professionals under the Italian Ministry of Health guidelines of 2 April 2025.
Enter as a healthcare professionalFrequently asked questions
Who may use HMS Oxy Evo?
How are the applied parts for oxygen and ozone connected?
Which gel should be used for the handpieces with electrodes?
Related devices
HMS Carbo
Information reserved for healthcare professionals
Healthcare professionals area →HMS Omnia
Information reserved for healthcare professionals
Healthcare professionals area →HMS Omnia Wave
Information reserved for healthcare professionals
Healthcare professionals area →Want to see it working in your treatment room?
On-site or online demo, quote and rental simulation within 48 hours.