HMS Omnia Wave
Electromedical device with electromagnetic and/or pneumatic shock waves and electroporation
- EM shock waves
- 1–16 Hz · 0.06–3.68 mJ/mm²
- PN shock waves
- 1–16 Hz · 0.1–3.0 bar
- Heads
- Radial 8/15/36 mm · Focal 15/20/39 mm
- Electroporation
- 120 V · 1–500 Hz
- Display
- 15" touch
- Weight
- 10 kg
Intended use
What it is for and who uses it
HMS OMNIA WAVE is a non-invasive active therapeutic electromedical device employing the following technologies: Shock waves to treat, in the osteoarticular field, delayed bone consolidation, plantar fasciitis, epicondylitis, calcific and non-calcific tendinitis, tendinopathies such as patellar and Achilles; in the field of venous and lymphatic microcirculation, Oedematous Fibrosclerotic Panniculopathy and Lipodystrophy. Electroporation for the transdermal inoculation of therapeutic active ingredients. In addition, HMS OMNIA WAVE employs the Shock Wave and Electroporation technologies also for aesthetic applications outside the scope of Directive 93/42/EEC as amended, namely respectively for body contouring and the transdermal delivery of cosmetic substances.
Qualified operators
The intended use of the appliance is that of a "medical device", used exclusively by personnel qualified in accordance with the law for applications in the fields of physiotherapy, rehabilitation, pain therapy and aesthetic medicine. Specific training is required for correct use of the product.
Treatments and areas
- Delayed bone consolidation
- Plantar fasciitis
- Epicondylitis
- Calcific and non-calcific tendinitis
- Patellar and Achilles tendinopathies
- Oedematous Fibrosclerotic Panniculopathy
- Lipodystrophy
- Transdermal inoculation of therapeutic active ingredients
Contraindications
- Shock waves: contraindicated in cases of severe decompensated heart disease; for wearers of active implants such as pacemakers, insulin pumps, pericardial stimulators, hearing aids or others; metal prostheses close to the treated area; pregnancy and breastfeeding; in subjects suffering from epilepsy; heart disease and blood circulation disorders (cardiac arrhythmia, capillary fragility, use of anticoagulant drugs, deep venous insufficiency, phlebitis, thrombopathies, haemorrhagic tendencies and disorders of the coagulation system with supporting drugs); osteoporosis; radial nerve compression; hyperthyroidism; inflammatory or neoplastic disorders; acute inflammation in the treatment area; polyneuropathies of diabetes mellitus; cancer; on skeletal tissue that is not fully developed. Cortisone and NSAIDs must not be administered in the period preceding the treatment.
- Electroporation: contraindicated for wearers of active implants such as pacemakers, insulin pumps, pericardial stimulators, hearing aids or others; for women who are pregnant and/or breastfeeding; for wearers of metal prostheses and osteosynthesis devices (including dental prostheses) close to the area to be treated; for subjects with ongoing inflammatory processes, skin lesions, neoplasms; do not treat subjects with sensitive skin, if necessary limit to very low output powers; during menstruation (unless prescribed by a physician); apply only to intact skin; in subjects with multiple sclerosis (owing to spastic paralysis and sensory disorders); in persons with Parkinson's disease (owing to rigidity from increased muscle tone and tremor); in the presence of febrile illness; in cases of epilepsy; in cases of thrombophlebitis; in cases of heart disease. Avoid applying the electrodes on the carotid sinuses, so as not to stimulate the vagovagal reflex, and around the head.
- In case of doubt, do not use without first consulting the opinion of the specialist physician.
List taken from the instructions for use. Read the warnings and instructions carefully before use.
Technology
HMS Omnia Wave is a non-invasive active therapeutic electromedical device that implements shock wave technologies in the fields of physiotherapy and aesthetic medicine, together with electroporation. It is offered in five configurations (REF HMS.O.W-EPR, -EP, -ER, -E, -P) combining electromagnetic and/or pneumatic shock waves, with dedicated handpieces; electroporation is included in all of them. The device is controlled by a 15-inch colour touch display and a footswitch; cooling is provided by a distilled water circuit with fill, vent and drain connections on the rear.
Electromagnetic shock waves
According to the manual, in physiotherapy shock waves are recognised in the treatment of certain osteoarticular system conditions such as tendinopathies, fasciitis and delayed bone consolidation: in addition to promoting revascularisation of the treated areas, they cause an intense stimulation that activates the small-diameter fibres projecting into the periaqueductal grey area, activating the serotonergic system that modulates the transmission of pain stimuli. In aesthetic medicine, accompanied by vacuum to focus the energy, they are described as a non-invasive procedure for age-related laxity of the connective tissue, in particular in the treatment of the blemishes of cellulite and lipodystrophies, stimulating microcirculation in the adipose tissue.
The HMS.MOU.E electromagnetic handpiece works with a frequency from 1 to 16 Hz (1 Hz steps) and an energy density from 0.06 to 3.68 mJ/mm² (0.01 mJ/mm² steps), with radial heads of 8, 15 and 36 mm and focal heads of 15 and 20 mm. Delivery is enabled by the footswitch and/or the button on the handpiece.
Pneumatic shock waves
Configurations HMS.O.W-EPR, -EP and -P include the HMS.MOU.P pneumatic handpiece, with a frequency from 1 to 16 Hz and a working pressure adjustable from 0.1 to 3.0 bar (0.1 bar steps); the focal head has a diameter of 39 mm. Parameters are set from the touch display and the power can be adjusted during treatment.
Electroporation
Applying a particular electrical pulse to a biological surface generates aqueous channels in the cell membrane, known as electropores: the phenomenon induces a significant transient increase in tissue permeability, which promotes the transcutaneous absorption of water-soluble active ingredients. In the intended use, electroporation is provided for the transdermal inoculation of therapeutic active ingredients and, in the aesthetic field outside the scope of Directive 93/42/EEC, for the transdermal delivery of cosmetic substances. The section delivers up to 120 V @ 1000 ohm with a frequency from 1 to 500 Hz; depth of action, body area, intensity (1–100 %) and time are set from the display. The equipment includes the HMS.MCEP applicator, the synthetic chamois electrode, the cable kit and the 40-60-100 elastic straps.
Handpieces and standard equipment
What comes in the box
Consumables: CE-marked gel for the shock wave handpieces, Distilled water for the cooling circuit, Active ingredients for electroporation
Technical specifications
Technical specifications
| Power supply | 110–230 V AC, 50–60 Hz |
|---|---|
| Max power consumption | 300 W |
| Electrical safety class | II BF |
| IP rating | IP40 |
| External protection | Protection fuses 2 × 3.15 A |
| Cooling | Distilled water hydraulic circuit (replacement every 15 days) |
| HMS.O.W-EPR and HMS.O.W-EP | EM shock wave handpiece + PN shock wave handpiece + electroporation |
|---|---|
| HMS.O.W-ER and HMS.O.W-E | EM shock wave handpiece + electroporation |
| HMS.O.W-P | PN shock wave handpiece + electroporation |
| Common equipment | Cable, footswitch, electroporation applicator, chamois electrode, cable kit, strap kit |
| Emission frequency | 1 ÷ 16 Hz software-selectable (1 Hz steps) |
|---|---|
| Energy density | 0.06 ÷ 3.68 mJ/mm² (0.01 mJ/mm² steps) |
| Head diameter | Radial 8 mm, 15 mm, 36 mm · Focal 15 mm, 20 mm |
| Delivery | Footswitch and/or button on the handpiece |
| Emission frequency | 1 ÷ 16 Hz software-selectable (1 Hz steps) |
|---|---|
| Working pressure | 0.1 ÷ 3.0 bar (0.1 bar steps) |
| Head diameter | Focal 39 mm |
| Max output voltage | 120 V @ 1000 ohm |
|---|---|
| Emission frequency | 1 ÷ 500 Hz |
| Intensity | 1 ÷ 100 % (1 % steps) |
| Parameters | Depth of action, body sector/area, intensity, time |
| Display | 15" colour touch screen |
|---|---|
| Dimensions (H × W × D) | 45 × 35 × 40 cm |
| Weight | 10 kg |
| Operating temperature | 0 °C ÷ +40 °C |
| Operating humidity | 30 % ÷ 75 % non-condensing |
| Operating pressure | 700 ÷ 1060 hPa |
| Storage and transport | −10 °C ÷ +70 °C · max humidity 95 % · 500 ÷ 1060 hPa |
| Output parameter tolerance | 20 % |
Output parameter tolerance: 20% (data from the instructions for use).
Compliance and certifications
Medical device · CE marking with Notified Body
- Notified Body
- 1370
- Manufacturer
- CTM Tech S.r.l.s., Terni (Italia)
- EUDAMED SRN
- IT-MF-000032025
- Quality system
- ISO 13485 · AQSCert
- IFU
- Rev. 01 · 2022-07-26
Documentation
Training and service
Training included at delivery, direct service from our Terni workshop, replacement unit in case of extended downtime, CEI EN 62353 electrical checks.
Also available for rental or leasing → options
Healthcare professionals area
HMS Omnia Wave – information for healthcare professionals
Content reserved for healthcare professionals under the Italian Ministry of Health guidelines of 2 April 2025.
Enter as a healthcare professionalFrequently asked questions
Which configurations are available?
Who may use HMS Omnia Wave?
What maintenance does the cooling circuit require?
Related devices
HMS Carbo
Information reserved for healthcare professionals
Healthcare professionals area →HMS Omnia
Information reserved for healthcare professionals
Healthcare professionals area →HMS Oxy Evo
Information reserved for healthcare professionals
Healthcare professionals area →Want to see it working in your treatment room?
On-site or online demo, quote and rental simulation within 48 hours.